文摘
选择性的分析性能,自动multianalyser奥林巴斯盟5031是实用性和评估评估四个多月八个月。评价是摘自传道的指导方针。20例程参数测量。此外,钠和钾测定附加火焰光度单位。双方之间的协议单位和8光度计比色皿的温度行为是令人满意的。不精确是非常好。within-run裁的大多数参数低于1.5%。天之间的不精确是低于5%,除了肌酸磷酸激酶(7.4%)。谷氨酸脱氢酶给4.0%到15.9%的不精确,,然而,更有可能是由于测量的低活动而不是分析器的错。指定值的恢复在12控制血清为14个测试在95%和105%之间。 Three of the remaining eight tests yielded recoveries with deviations between 10% and 18% (alanine aminotransferase, aspartate aminotransferase and bilirubin). No drift effects were observed and neither a sample carry-over nor a reagent carry-over were detected. Most tests were linear over a very wide range. Only afew tests (mainly lipase and glutamate dehydrogenase) required measurement repetitions with diluted samples. The correlation with routine instruments and tests was close. However, corrections were necessary for 14 of the 22 tests. This was not due to the performance of the analyser but, rather, to the different methodologies of compared tests, or different working temperatures on the comparison instruments, or a lack of accuracy for some of the AU tests.