TY - JOUR A2 - Pillar, Giora AU - Stepnowsky, Carl AU - Zamora, Tania AU - Barker, Robert AU - Liu, Lin AU - Sarmiento, Kathleen PY - 2013 DA - 2013/08/25 TI -正压通气装置测量的呼吸暂停和低呼吸的准确性:SP - 314589 VL - 2013 AB -改进的数据传输技术有助于从家庭环境中的正压通气(PAP)设备收集数据。尽管临床医生的治疗决策越来越依赖PAP装置对呼吸事件的自动评分,但很少有研究专门检查持续使用PAP的家庭环境中自动评分呼吸事件的准确性。在“PAP疗效”研究中,参与者同时佩戴一套Embletta睡眠系统(Embla, Inc., Broomfield, CO),该系统通过专用电缆直接连接到ResMed AutoSet S8 (ResMed, Inc., San Diego, CA)。平均pap评分的呼吸暂停低通气指数(AHI)为14.2±11.8(中位数:11.7;范围:3.9-46.3),平均手动评分AHI为9.4±10.2(中位数:7.7;范围:1.2 - -39.3)。计算各平均指标之间的比值。pap评分的HI是手工评分的2.0倍。pa评分的AHI是人工评分的1.5倍,AI是人工评分的1.04倍。 In this sample, PAP-scored HI was on average double the manual-scored HI. Given the importance of PAP efficacy data in tracking treatment progress, it is important to recognize the possible bias of PAP algorithms in overreporting hypopneas. The most likely cause of this discrepancy is the use of desaturations in manual hypopnea scoring. SN - 2090-3545 UR - https://doi.org/10.1155/2013/314589 DO - 10.1155/2013/314589 JF - Sleep Disorders PB - Hindawi Publishing Corporation KW - ER -